dc.description.abstract | Fluoroscopy-guided interventional procedures have gained prominence in diagnostic and
therapeutic settings due to their minimally invasive nature. Among these procedures,
Coronary Angiogram (CAG) stands out as one of the most commonly performed cardiac
interventions. To minimize patient radiation exposure, the International Commission
of Radiation Protection (ICRP) recommends to establish Diagnostic Reference Levels
(DRLs) nationally or regionally. This study aims to establish an institutional DRL for
CAG procedures at a selected private hospital in Colombo, Sri Lanka. The research
was conducted in the cardiac catheterization laboratory, focusing on 201 CAG procedures
performed using two separate C-arm machines. For machine No: 01- PHILIPS ALLURA
clarity FD-10, the median cumulative DAP was 9060.50 mGy.cm2 while it was 7498.00
mGy.cm2 for machine No: 02- PHILIPS AZURION 7 B20/15. These median values were
considered as their respective DRLs for CAG procedures in each machine according to
the ICRP recommendation. In addition to establishing the DRLs, the study identi ed
signi cant correlations between patient e ective dose and several factors, including patient
weight, uoroscopy time, and the number of images taken. In conclusion, this
study established institutional DRLs for CAG procedures and highlights the importance of
optimizing radiation dose to achieve better patient outcomes. Established DRL values for
DAP were quite low compared with previous studies. Therefore, this study recommends to
develop national DRLs for interventional procedures to optimize the radiation exposure
level. | en_US |
dc.language.iso | en | en_US |
dc.subject | Catheterization laboratory, | en_US |
dc.subject | CAG, | en_US |
dc.subject | Cumulative dose area product, | en_US |
dc.subject | Dose reference level, | en_US |
dc.subject | Radiation dose optimization | en_US |
dc.title | Development of Institutional Reference Level of Radiation Dose for Cardiac Catheterization Procedures | en_US |
dc.type | Article Abstract | en_US |
dc.identifier.faculty | Faculty of Health Sciences | en_US |
dc.identifier.journal | KDU IRC | en_US |